ASL Environmental Sciences Inspection, Survey & Testing
Calibration testing tank
Two-part calibrate-and-verify process for AZFP echosounders in ASL's freshwater test tank.
Description
ASL calibrates and verifies AZFP scientific echosounders in its freshwater test tank. Calibrated instruments let researchers quantify observed backscatter on an absolute scale and compare backscatter across frequencies (dB differencing) for classification.
Key features
- Calibrate: a calibrated source, then a calibrated hydrophone, is aligned with the transducer at one meter range to characterize the receiver and transmitter
- Verify: a tungsten-carbide sphere is moved through each beam in a large freshwater tank
- Pass criterion: measured target strength within +/- 1 dB of theoretical
Applications
Scientific studies of spatial and temporal changes in acoustic backscatter from zooplankton, fish and suspended material.